ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM, ACCESSORY, OZIL TORSIONAL HANDPIECE; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065751761
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Corneal Decompensation (1790); Eye Burn (2523)
Event Date 03/31/2022
Event Type  Injury  
Event Description
A physician reported that during an ophthalmic surgery the patient experienced thermal burn and corneal endothelial decreases.The surgery was completed.Additional information has been requested none received till date.
 
Manufacturer Narrative
A product sample has been received by the manufacturer and it is awaiting evaluation.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
The phacoemulsification (phaco) handpiece was received, and a visual assessment of the returned sample revealed no visual nonconformities.The returned phaco handpiece sample was connected to a calibrated ophthalmic vision system.The phaco handpiece tuned successfully and completed a five minute burn-in test with the system set at 100% ultrasonic and torsional power.The phaco handpiece was connected to dynamic tuning fixture (dtf) for stroke length testing on the longitudinal and torsional movements which found the phaco handpiece to meet product specifications.A phaco handpiece manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this lot/batch/serial number was performed.No similar complaints were reported for the product lot/batch/serial under investigation.The product under investigation is not a serviceable device.Therefore, a service record review was not performed.The phaco handpiece was found to meet specifications; therefore, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION VISION SYSTEM, ACCESSORY, OZIL TORSIONAL HANDPIECE
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14186558
MDR Text Key289866248
Report Number2028159-2022-00552
Device Sequence Number1
Product Code HQC
UDI-Device Identifier00380657517619
UDI-Public00380657517619
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K121555
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number8065751761
Device Lot Number128HV5
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/31/2022
Initial Date FDA Received04/22/2022
Supplement Dates Manufacturer Received07/07/2022
Supplement Dates FDA Received08/01/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/10/2019
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
CENTURION VISION SYSTEM
Patient Outcome(s) Other;
-
-