13 depth electrodes were involved in the case.Return inspection confirmed signal loss on some or all contacts of each electrode.Quantity, catalog #, udi, mfg.Date 2, sd06r-sp10x-000, (b)(4), 01/04/2022; 1, sd08r-sp05x-000, (b)(4), 01/28/2022; 1, sd08r-sp05x-000, (b)(4), 12/04/2021; 1, sd10r-sp05x-000, (b)(4), 12/16/2021; 5, sd10r-sp05x-000, (b)(4), 02/18/2022; 1, sd12r-sp05x-000, (b)(4), 01/05/2022; 1, sd14r-sp05x-000, (b)(4), 12/06/2021; 1, sd14r-sp05x-000, (b)(4), 01/12/2022.2 subdural electrodes were also involved in the case.Reported under mdr 2183456-2022-00008.
|