MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; IMPLANTED SUBCORTICAL ELECTRICAL STIMULATOR (M Back to Search Results
Model Number 37612
Device Problems Improper or Incorrect Procedure or Method (2017); Battery Problem (2885)
Patient Problem Insufficient Information (4580)
Event Date 04/19/2019
Event Type  malfunction  
Event Description
Information was received from the manufacturer¿s representative (rep) who reported therapy didn't work well for the patient, so they stopped using the system.The system was likely in overdischarge, so the rep needed assistance to start the physician recharge mode (pmr) as the patient needed an mri as the patient was going to have a surgical procedure to address their dystonia symptoms rather than using the system for symptom relief.Following initiation of the pmr session the session stopped, and a power on reset (por) was seen.The rep then reported the ins was charging normally and the por was cleared allowing the device to be interrogated.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
IMPLANTED SUBCORTICAL ELECTRICAL STIMULATOR (M
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key14192026
MDR Text Key289956310
Report Number3004209178-2022-05168
Device Sequence Number1
Product Code MRU
UDI-Device Identifier00643169864238
UDI-Public00643169864238
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H020007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/14/2020
Device Model Number37612
Device Catalogue Number37612
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/18/2022
Initial Date FDA Received04/25/2022
Date Device Manufactured03/14/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age40 YR
Patient SexFemale
-
-