The customer reported during a therapeutic ureteroscopy (urs) to remove a stone, the subject device bending section was heavily damaged due to a shot with a non-olympus laser.A piece of the non-olympus laser fiber broke off and was left in the patient.The intended procedure was cancelled because the stone was too large to be removed.There was no clinically relevant extension of the procedure and no harm to the patient occurred.Another procedure, percutaneous nephrolithotomy, was performed to remove the large stone and the non-olympus device fragment.No harm occurred, the patient is doing well.
|
This supplemental report is being submitted to provide the results of the legal manufacturer¿s investigation and the results of the device history records (dhr) review.The dhr for this device were reviewed and all records indicated the product was manufactured according to all applicable procedures and met final product release criteria.No abnormalities were found.A definitive root cause was not identified.Based on the available information, the legal manufacturer determined the probable cause of the failure is likely due to physical stress by the user.Ifu (operation manual) states on cautions of inserting laser probe as below: ¿4.4 using the laser system: caution do not insert the laser probe into the instrument channel while applying an excessive force or bending the bending section.The instrument channel and/or laser probe may be damaged.¿ ifu (operation manual) states on handling of insertion section and bending section as below: ¿important information ¿ please read before use.¦warnings and cautions: warning.Do not strike, hit, or drop the endoscope¿s distal end, insertion tube, bending section, or control section.Also, do not bend, pull, or twist the endoscope¿s distal end, insertion tube, bending section, control section, or light guide cable with excessive force.The endoscope may be damaged and could cause patient injury, burns, bleeding, and/or perforations.It could also cause parts of the endoscope to fall off inside the patient.¿ ifu (operation manual) states on the inspection prior to user for bending section as below: ¿3.3 inspection of the endoscope.¦inspection of the endoscope.11 inspect the bending section for bends, twist, or other irregularities while the bending section remains straight.12 inspect the bending section for abnormal bending shape, or other irregularities.¿ ifu (operation manual) states on caution for angulation manipulation as below: ¿4.1 precautions: caution.Do not operate the angulation control lever with excessive force in a narrow space to the opposite direction from the bending direction while the distal end of the endoscope is not moved.The bending section may be damaged.Check the tip position of the endoscope and the shape of the bending section using fluoroscopy, etc.Do not insert the insertion tube with excessive force and twist.¿ additional ¿instructions for safe use¿ also states on angulation manipulation so as not to cause damage to bending section as below: ¿4.1 precautions: caution.Do not operate the angulation control lever with excessive force in a narrow space to the opposite direction from the bending direction while the distal end of the endoscope is not moved.The bending section may be damaged.Check the tip position of the endoscope and the shape of the bending section using fluoroscopy, etc.Do not insert the insertion tube with excessive force and twist.¿ olympus will continue to monitor field performance for this device.
|