Model Number max 2
Device Problem Use of Device Problem (1670)
Patient Problem Burn(s) (1757)
Event Date 03/27/2022
Event Type  malfunction  
Event Description
The user facility reported an employee experienced a burn on their hands while handling items that were processed in a v-pro max 2 sterilizer.The employee sought medical treatment at the facility's er and was sent home.
 
Manufacturer Narrative
A steris service technician arrived onsite following the reported event to inspect the v-pro max 2 sterilizer and found the unit to be operating properly.No repairs were required and the unit was returned to service.Through follow-up with user facility personnel, the technician was informed that the employee was not wearing proper ppe, specifically gloves as stated in the operator manual.The v-pro max 2 sterilizer operator manual states (6-27), "steris recommends (in accordance with ansi/aami st58, 2013) wearing chemical-resistant gloves when using the sterilization unit." the operator manual further states (1-2), "danger - chemical injury hazard: any visible liquids in the chamber or on the load must be treated as concentrated hydrogen peroxide.Observe all hydrogen peroxide handling precautions.When handling hydrogen peroxide, wear appropriate personal protective equipment." additionally, user facility personnel should ensure all instruments are properly dry prior to placement in the v-pro max 2 sterilizer.The v-pro max 2 operator manual states (a-1), "dry all items thoroughly.Ensure all moisture is removed from all internal parts (including lumens).If not, residual hydrogen peroxide may remain at cycle completion and/or a cycle abort occurs.Only dry items are to be placed in sterilization unit." the technician counseled the user facility on the proper use and operation of the v-pro max 2 sterilizer, specifically wearing proper ppe and properly drying instruments before processing.No additional issues have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V-PRO MAX 2 STERILIZER
Type of Device
STERILIZER
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX  67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX   67190
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key14196680
MDR Text Key291498237
Report Number3005899764-2022-00016
Device Sequence Number1
Product Code MLR
UDI-Device Identifier00724995158354
UDI-Public00724995158354
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Numbermax 2
Device Catalogue NumberVP50002101
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/27/2022
Initial Date FDA Received04/25/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/28/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-