Model Number 2ACH20
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Low Blood Pressure/ Hypotension (1914); Hypoxia (1918); Pericardial Effusion (3271)
Event Date 04/11/2022
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryo ablation procedure, the patient's oxygen saturation levels dropped.The patient was supported and medications were reversed.  the blood pressure began to decrease. intracardiac echocardiography (ice) was performed and an effusion was observed. pericardiocentesis was performed.The patient was stabilized and intubated.The last ablations were completed. the case was completed with cryo.The patient was admitted to the intensive care unit (icu) for monitoring.The patient's hospitalization was extended one day.The physician does not know what caused the effusion. no further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the data files were returned and analyzed.The data files showed at least 17 applications were performed with a non-returned balloon catheter without any issue on the date of the event.In conclusion, the clinical issues (hypotension and pericardial effusion) occurred during the procedure.There is no indication of relation of the adverse events to the performance or a malfunction of the product.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACHIEVE ADVANCE MAPPING CATHETER
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECT
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key14197176
MDR Text Key290071819
Report Number2182208-2022-01318
Device Sequence Number1
Product Code DRF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162892
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number2ACH20
Device Catalogue Number2ACH20
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/21/2022
Initial Date FDA Received04/25/2022
Supplement Dates Manufacturer Received04/27/2022
Supplement Dates FDA Received05/23/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
AFAPRO28 BALLOON CATHETER, 4FC12 SHEATH
Patient Outcome(s) Life Threatening; Required Intervention; Hospitalization;
Patient SexMale
Patient Weight84 KG
-
-