There was an allegation of questionable elecsys anti-tpo results for 3 patients on a cobas e 801 analytical unit, serial number (b)(4).Patient 1: the initial result was 80.1 iu/ml (primary tube).On (b)(6) 2022, the repeated result was 15.0 iu/ml (from the same primary tube).Patient 2: on (b)(6) 2022, the initial result was 70.1 iu/ml (primary tube).On (b)(6) 2022, the repeated result was 9 iu/ml (from the same primary tube).Patient 3: on (b)(6) 2022, the initial result was 66 iu/ml.On (b)(6) 2022, the repeated result was <34 iu/ml.The repeated results were performed on another e801 module.The results were reported outside of the laboratory.The repeated results were believed to be correct.
|