Model Number 101/870/080CZ
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/23/2022
Event Type  Injury  
Event Description
It was reported that during the use of the product, leakage of air from the product was observed.No patient injury.
 
Manufacturer Narrative
H10: device evaluation: one used decontaminated sample 10009163-004 8mm deht blu suctionaid sub-assy 1/ea* was returned for investigation without its original packaging.Under visual inspection we noticed a tear in cuff (see photo and video of defect).During manufacturing process the devices are 100% inflation tested, which includes inflating each device cuff and leaving for a 12 hour period.Reductions in pressure over this time are considered a failure and the device would be rejected.The inflation test was repeated on the received sample.It was found that it is not even possible to inflate cuff as it leaks so badly.Due to fact that cuff leak was observed after placement it is the most probable that reported failure occurred during tracheostomy procedure or during use.The cause of the reported problem was traced to the user actions.A dhr review was performed subsequent to the manufacturing of the device and prior to its release.No problems or issues were identified during this dhr review.
 
Manufacturer Narrative
This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX BLUE LINE ULTRA SUCTIONAID
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
NULL
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
2-12-12, kitanagase-omotemachi
minneapolis, MN 55442
MDR Report Key14208530
MDR Text Key290082749
Report Number3012307300-2022-06963
Device Sequence Number1
Product Code BTO
UDI-Device Identifier15019517076226
UDI-Public15019517076226
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K173384
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 05/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number101/870/080CZ
Device Catalogue Number101/870/080CZ
Device Lot Number4070522
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/04/2022
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/29/2022
Initial Date FDA Received04/26/2022
Supplement Dates Manufacturer Received05/17/2022
05/06/2023
Supplement Dates FDA Received06/02/2022
05/11/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/22/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-