Model Number M00522610 |
Device Problem
Activation, Positioning or Separation Problem (2906)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/21/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Initial reporter's address: (b)(6).Medical device code (b)(4).The device has not been received for analysis.Upon receipt and completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
|
Event Description
|
It was reported to boston scientific corporation that a resolution clip device was used in the colon during an endoscopic mucosal resection (emr) procedure performed on (b)(6) 2022.During the procedure, the clip was able to grasp and lock onto tissue; however, the clip was unable to release from the catheter to deploy.The procedure was completed with another resolution clip device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
|
Manufacturer Narrative
|
Initial reporter's address: (b)(6).Medical device code a15 captures the reportable event of clip unable to release from the catheter.Block h10: investigation results the returned resolution clip device was analyzed, and a visual evaluation noted that the device was returned without the clip assembly and with evidence of full deployment.The device was returned without the over-sheath.Microscopic examination was performed and it was found that the bushing had hit marks on the hooks.Dimensional examination was performed on the bushing outer diameter, and it was found to be within specification.A dimensional analysis was also performed between the hooks of the bushing and both sides were noted to be within specification.No other problems with the device were noted.The reported event of clip unable to release from the catheter was not confirmed.Investigation found that the bushing had hit marks on the hooks, this is likely due to the interaction between the yoke and the capsule, in order to deploy the clip.The device was returned without the clip assembly and it is important to mention that the presence of this component is very important to the investigation in order to analyze any failure that could contribute with the problem faced by the physician.The investigation findings and all information available do not lead to a clear conclusion about the cause of the reported adverse event.Therefore, the most probable root cause is cause not established.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
|
|
Event Description
|
It was reported to boston scientific corporation that a resolution clip device was used in the colon during an endoscopic mucosal resection (emr) procedure performed on (b)(6), 2022.During the procedure, the clip was able to grasp and lock onto tissue; however, the clip was unable to release from the catheter to deploy.The procedure was completed with another resolution clip device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
|