SYNTHES GMBH 3.2MM DRILL BIT/QC/145MM; INSTRUMENT, SURGICAL, ORTHOPEDIC, PNEUMATIC POWERED & ACCESSORY/ATTACHMENT
|
Back to Search Results |
|
Catalog Number 310.310 |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/01/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product complaint # (b)(4).Complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.A manufacturing record evaluation was performed for the finished article lot and no non-conformance's were identified.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Device report from synthes reports an event in (b)(6) as follows: it was reported that the drill bit breaks in surgery.This complaint involves one (1) device drill bit ø3.2 l145/120 2flute.This is report 1 of 1 for complaint (b)(4).
|
|
Event Description
|
It was noted no fragments were left in the patient and the surgical time was not extended.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10: additional narrative: b5.H3, h6: a product investigation was completed: visual analysis of the returned sample revealed that the distal tip of drill bit has broken; the broken fragment was not returned.No other problems were reported.A dimensional inspection was not performed due to post manufacturing damage.The current and manufactured to drawing was reviewed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the drill bit would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11: corrected data: d9, h4.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d2: additional procodes: hwe, gfa, gff d7 h6: codes added for both completed photo investigation and pending physical investigation.H6: part: 310.310 lot: 73p7049 manufacturing site: (b)(4) release to warehouse date: 13 october 2020 a manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.The photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the device from the received image.Visual analysis of the photo revealed that the drill bit ø3.2 l145/120 2flute appears to be broken from the tip; fragment is not visible in the evidence provided.Additionally, the remaining section of the tip is deformed.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for the drill bit ø3.2 l145/120 2flute.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D1, d2, d9 h4.
|
|
Search Alerts/Recalls
|
|
|