It was reported that the patient had not benefitted from vns and had their device explanted.Explanted products were returned for product analysis.Product analysis was completed on the lead.Continuity checks of the returned lead portion were performed during the functional analysis, and no discontinuities were identified.The condition of the returned lead portion is consistent with conditions that typically exist following an explant procedure.Based on the findings in the product analysis lab, other than typical wear and explant related observations, no anomalies were identified in the returned lead portion.Note that since a portion of the lead assembly (body) including the electrode array section was not returned for analysis, an evaluation and resulting commentary cannot be made on that portion of the lead.Abrasions were noted in multiple locations, and the lead assembly has dried remnants of what appear to have once been body fluids inside the outer and inner silicone tubes.This was noted to possibly be due to wear.
|