Model Number 0620040600
Device Problems Increase in Pressure (1491); Pressure Problem (3012)
Patient Problems Pneumothorax (2012); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 04/05/2022
Event Type  malfunction  
Event Description
It was reported that patient had suffered a collapsed lung due to overpressure.
 
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event Description
It was reported that patient had suffered a collapsed lung due to overpressure.
 
Manufacturer Narrative
Alleged failure: (b)(4), biomed, the pressure is fluctuating dramatically and not consistent, po # (b)(4).Update: the statement made to me on the phone was "the pressure shot up to 21, so high that it collapsed the patients lungs".They then put the unit in pediatric mode and have no further issues.The failure alleged in the complaint record was not confirmed/duplicated during the product investigation.The probable root cause/s could be an incorrect orientation of the o-ring on the tubeset adapter and/or the venting system being off which would cause the overpressure to not be relieved.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Manufacture date is not known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG; PNEUMO SURE HIGH FLOW INSUFFLATOR
Type of Device
INSUFFLATOR, LAPAROSCOPIC
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
chester rebugio
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key14240411
MDR Text Key290365496
Report Number0002936485-2022-00205
Device Sequence Number1
Product Code HIF
UDI-Device Identifier07613327063219
UDI-Public07613327063219
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063367
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0620040600
Device Catalogue Number0620040600
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/05/2022
Initial Date FDA Received04/28/2022
Supplement Dates Manufacturer Received04/05/2022
Supplement Dates FDA Received06/09/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-