• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPORT DUO M.R.I. IMPLANTABLE PORT, CHRONOFLEX DUAL-LUMEN, 9.5F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPORT DUO M.R.I. IMPLANTABLE PORT, CHRONOFLEX DUAL-LUMEN, 9.5F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Model Number 1829500
Device Problems Fluid/Blood Leak (1250); Material Protrusion/Extrusion (2979)
Patient Problems Extravasation (1842); Infiltration into Tissue (1931)
Event Date 03/28/2022
Event Type  Injury  
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiration date: 05/2022).
 
Event Description
It was reported that approximately six months and fifteen days post port implant procedure, the port septum allegedly leaked.It was further reported that a pop was heard when flushing and caused infiltration of fluid under the skin.Reportedly the port was removed and replaced with another port.There was no reported patient injury.
 
Event Description
It was reported that approximately six months post port implant procedure, the port septum allegedly dislodged and leaked.It was further reported that a pop was heard when flushing and caused infiltration of fluid under the skin.Reportedly the port was removed and replaced with another port.The current status of the patient is unknown.
 
Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one powerport duo mri implantable port attached to a catheter was returned for evaluation.Visual, microscopic visual and functional evaluation were performed on the returned device.The investigation is confirmed for the reported material protrusion, infiltration and fluid leak issue as one of the two port septa(right) was noted to be dislodged from the port body and upon infusing the port, leak was noted from the partially dislodged septum.Further during functional evaluation mandrel test was performed were both the port stems passed the test as the black mark on the mandrel was fully inserted within each lumen.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiration date: 05/2022).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPORT DUO M.R.I. IMPLANTABLE PORT, CHRONOFLEX DUAL-LUMEN, 9.5F
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key14249616
MDR Text Key290370680
Report Number3006260740-2022-01556
Device Sequence Number1
Product Code LJT
UDI-Device Identifier00801741027185
UDI-Public(01)00801741027185
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090512
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1829500
Device Catalogue Number1829500
Device Lot NumberREFQ2563
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/19/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/04/2022
Initial Date FDA Received04/29/2022
Supplement Dates Manufacturer Received06/28/2022
Supplement Dates FDA Received07/01/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-