H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one powerport duo mri implantable port attached to a catheter was returned for evaluation.Visual, microscopic visual and functional evaluation were performed on the returned device.The investigation is confirmed for the reported material protrusion, infiltration and fluid leak issue as one of the two port septa(right) was noted to be dislodged from the port body and upon infusing the port, leak was noted from the partially dislodged septum.Further during functional evaluation mandrel test was performed were both the port stems passed the test as the black mark on the mandrel was fully inserted within each lumen.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiration date: 05/2022).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|