• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SP*2 FEMORAL IMPACTOR; KNEE INSTRUMENT : IMPACTORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SP*2 FEMORAL IMPACTOR; KNEE INSTRUMENT : IMPACTORS Back to Search Results
Model Number 96-6180
Device Problems Break (1069); Naturally Worn (2988); Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/20/2022
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Instruments are broken due to wear and tear.No patient involvement.
 
Event Description
Additional information received indicated that the femoral impactor was loose and no longer able to be screwed back tight, it was in two parts.No items had broken into more than one piece.
 
Manufacturer Narrative
Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: b5.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: e4, h6 (medical device problem code).
 
Manufacturer Narrative
Product complaint (b)(4).Investigation summary: no device was received for examination.Therefore, the reported event could not be confirmed.Depuy considers, the investigation closed.Should additional information be received, the information will be reviewed.And the investigation will be re-opened as necessary.Device history lot: a manufacturing record evaluation (mre) was not possible, because the required lot code was not provided.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
A.Please provide the lot no of the following products: *sp*2 femoral impactor (966180) sp2 tibial radel impactor (258111000).Pfcsig postlip trial 8mm sz5 (961260).Pfcsigma ins trl stabl 8mm sz4 (961450).I¿ve checked alll of these components, there are no visible lot numbers which i can see, b.Which part of the instruments were broke? sp*2 femoral impactor (966180) , this is loose , and no longer able to be screwed back tight, (it¿s in two parts).Sp2 tibial radel impactor (258111000), this has split, but is still intact, npbut not able to be used again on a patient.Pfcsig postlip trial 8mm sz5 (961260) this is split, at the back of the trial, pfcsigma ins trl stabl 8mm sz4 (961450) another split trial, in one piece , at the back where it sits into a trial tibial tray.C.Did they broke into 2 or more pieces? no items had broken into more than one piece.Please also provide us with an update with regards to the product return.Have you already returned the product for analysis? (if yes, please confirm the date shipped and the courier tracking number).If the hospital has not or will not release the product until a future date please confirm this and what the future date is.If the product is no longer available please advise.The hospital has discarded the items, they are no longer available to return to the analysis lab.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
KNEE INSTRUMENT : IMPACTORS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key14250235
MDR Text Key290380304
Report Number1818910-2022-07860
Device Sequence Number1
Product Code HWA
UDI-Device Identifier10603295247203
UDI-Public10603295247203
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date 04/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number96-6180
Device Catalogue Number966180
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/20/2022
Initial Date FDA Received04/29/2022
Supplement Dates Manufacturer Received05/05/2022
05/05/2022
05/30/2022
Supplement Dates FDA Received05/07/2022
05/24/2022
05/30/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-