MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION RESOLUTION CLIP; CLIP, HEMOSTATIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION RESOLUTION CLIP; CLIP, HEMOSTATIC Back to Search Results
Model Number M00522610
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/22/2022
Event Type  malfunction  
Manufacturer Narrative
Initial reporter address: (b)(6).(b)(4).The device has not been received for analysis.Upon receipt and completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation that a resolution clip device was used in the colon during an endoscopic mucosal resection (emr) procedure performed on (b)(6) 2022.During the procedure, the clip was able to grasp and lock onto tissue; however, the clip was unable to release from the catheter to deploy.The procedure was completed with another resolution clip device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Manufacturer Narrative
Block e1: initial reporter address: (b)(6).Block h6: medical device problem code a15 captures the reportable event of clip unable to release from catheter.Block h10: investigation results: the returned resolution clip device was analyzed, and a visual evaluation noted that the device was returned with the clip assembly attached.The device was returned without the over-sheath.Microscopic examination was performed and it was found that there was no activation performed on the clip assembly.Functional evaluation was performed and the clip assembly can open its arms within specification under the tortuous path.Additionally, the clip assembly was able to perform the first activation and could be released from the catheter without problem.No other problems with the device were noted.The reported event of clip unable to release from the catheter was not confirmed as the device was returned with the clip assembly still attached and the clip was able to release from the catheter without problem.No problems were found that could have reduced the performance of the device during the procedure.The returned device review showed no evidence of either the alleged(s) or any defect which could have contributed to the event.Therefore, the most probable root cause for this complaint is no problem detected.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
 
Event Description
It was reported to boston scientific corporation that a resolution clip device was used in the colon during an endoscopic mucosal resection (emr) procedure performed on (b)(6) 2022.During the procedure, the clip was able to grasp and lock onto tissue; however, the clip was unable to release from the catheter to deploy.The procedure was completed with another resolution clip device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTION CLIP
Type of Device
CLIP, HEMOSTATIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
2546 calle primera
propark, coyol
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key14250382
MDR Text Key290452906
Report Number3005099803-2022-02227
Device Sequence Number1
Product Code MCH
UDI-Device Identifier08714729504795
UDI-Public08714729504795
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K040148
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00522610
Device Catalogue Number2261
Device Lot Number0028212062
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/02/2022
Initial Date FDA Received04/29/2022
Supplement Dates Manufacturer Received08/05/2022
Supplement Dates FDA Received08/31/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/15/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age56 YR
Patient SexFemale
Patient Weight50 KG
-
-