• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIOMED GMBH & CO. MEDIZINTECHNIK KG LIFESTENT 5F VASCULAR STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIOMED GMBH & CO. MEDIZINTECHNIK KG LIFESTENT 5F VASCULAR STENT Back to Search Results
Catalog Number 5F060403C
Device Problem Positioning Problem (3009)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/14/2022
Event Type  malfunction  
Manufacturer Narrative
The catalog number identified has not been cleared in the us, but is similar to the lifestent 5f vascular stent system products that are cleared in the us.The 510 k number and pro code for the lifestent 5f vascular stent system products are identified.As the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiry date 02/2022).
 
Event Description
It was reported that during a stent placement procedure of a chronic total occlusion in the superficial femoral artery via a contralateral approach, the delivery system was allegedly unable to cross the lesion.There was no reported patient injury.
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us, but is similar to the lifestent 5f vascular stent system products that are cleared in the us.The 510 k number and pro code for the lifestent 5f vascular stent system products are identified in d2 and g4.H10: manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the sample was not returned to the manufacturer for evaluation and images were not provided for review.Based on the information available the investigation is inconclusive.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling for this product the potential issue was found addressed.The instructions for use state: 'if resistance is met during stent system introduction, the stent system should be withdrawn and another stent system should be used.' in regards to pta the instructions for use state: 'pre-dilatation of the lesion with a balloon dilatation catheter is recommended.' the instructions for use further state: 'examine the stent system for any damage.If it is suspected that the sterility or performance of the stent system has been compromised, the device should not be used.' , and 'do not use the device after the ¿use by¿ date specified on the label.' h10: d4 (expiry date 02/2022), g3.H11: g1, h6 (method).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that during a stent placement procedure of a chronic total occlusion in the superficial femoral artery via a contralateral approach, the delivery system was allegedly unable to cross the lesion.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFESTENT 5F VASCULAR STENT
Type of Device
VASCULAR STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM   76227
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key14284391
MDR Text Key290787138
Report Number9681442-2022-00140
Device Sequence Number1
Product Code NIP
UDI-Device Identifier00801741120060
UDI-Public(01)00801741120060
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P070014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number5F060403C
Device Lot NumberANEP3821
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/14/2022
Initial Date FDA Received05/04/2022
Supplement Dates Manufacturer Received05/10/2022
Supplement Dates FDA Received05/13/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexFemale
-
-