Catalog Number 8065751613
Device Problem Detachment of Device or Device Component (2907)
Patient Problem Retinal Detachment (2047)
Event Date 04/19/2022
Event Type  Injury  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A physician reported at the time of removing the vitrectomy cutter from the eye, the trocar came out with the cutter because it was very tight and caused retinal detachment during procedure.Procedure was completed with the same set-up.Sample was not available.
 
Manufacturer Narrative
A sample was not received at the manufacturing site for evaluation for the report of the trocar came out with the cutter causing retinal detachment; therefore, the condition of the product could not be verified.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.A sample was not received at the manufacturing site and the device history record review of the lot number provided indicated the product was processed and released according to the product¿s acceptable criteria, therefore; the root cause for the customer complaint issue cannot be determined.The exact root cause for this complaint is unknown, therefore, specific action with regards to this complaint cannot be taken.All trocar assemblies are 100% inspected for gauge size.Any non-conformance's found are removed from the lot and scrapped.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14285869
MDR Text Key290763498
Report Number1644019-2022-00330
Device Sequence Number1
Product Code LRO
UDI-Device Identifier00380657516131
UDI-Public00380657516131
Combination Product (y/n)N
Reporter Country CodeGR
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/28/2023
Device Catalogue Number8065751613
Device Lot Number2465851H
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/21/2022
Initial Date FDA Received05/05/2022
Supplement Dates Manufacturer Received07/29/2022
Supplement Dates FDA Received08/05/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/26/2021
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CONSTELLATION VISION SYSTEM
Patient Outcome(s) Other;
-
-