MAUDE Adverse Event Report: CONAIR LLC. CONAIR; HEATING PAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONAIR LLC. CONAIR; HEATING PAD Back to Search Results
Model Number HP01
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Burn(s) (1757)
Event Date 03/21/2022
Event Type  Injury  
Manufacturer Narrative
05/04/2022 - the device was returned to the manufacturer and is currently under investigation.
 
Event Description
(b)(6) 2022 - the consumer claims to have received a burn on her leg while in use of the product.The product was returned on 5/6/2022 and is currently being investigated.
 
Manufacturer Narrative
05/04/2022 - the device was returned to the manufacturer and is currently under investigation.6/2/2022 - the device was evaluated by the manufacturer.Below is the manufacturers narrative: manufacturers narrative: product was returned in good working condition.4/06/2022.Unit was run per ul130 temp set up.Temperature never exceeded 55c ( 131 f).Ul's limit is 70 c and conairs limit is 60c.We are under both the safety lab's and conairs temperature limits.Unit is not longer manufactured at this vendor.New construction was implemented in 2018.This construction is no longer manufactured.
 
Event Description
03/22/2022 - the consumer claims to have received a burn on her leg while in use of the product.The product was returned on 5/6/2022 and is currently being investigated.
 
Event Description
03/22/2022 - the consumer claims to have received a burn on her leg while in use of the product.The product was returned on 5/6/2022 and is currently being investigated.
 
Manufacturer Narrative
05/04/2022 - the device was returned to the manufacturer and is currently under investigation.6/2/2022 - the device was evaluated by the manufacturer.Below is the manufacturers narrative: 07/06/2022 - the was confusion between different complaints.Therefore the original manufacturer's narrative date 6/2/2022 was incorrect.Below is the corrected manufacturers narrative manufacturers narrative: unit was run per ul130 temp set up.Temperature never exceeded 55c ( 131 f).Ul's limit is 70 c and conairs limit is 60c.We are under both the safety lab's and conairs temperature limits.Unit performed as expected.Temperatures were lowered from competitive and safety labs limits to reduce burns.Consumer needs to be more diligent is checking skin prior to burns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONAIR
Type of Device
HEATING PAD
Manufacturer (Section D)
CONAIR LLC.
1 cummings point rd.
stamford CT 06902
Manufacturer Contact
1 cummings point rd.,
stamford, CT 06902
MDR Report Key14285877
MDR Text Key290771609
Report Number1222304-2022-00010
Device Sequence Number1
Product Code IRT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 03/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model NumberHP01
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/06/2022
Initial Date FDA Received05/05/2022
Supplement Dates Manufacturer Received04/06/2022
04/06/2022
Supplement Dates FDA Received06/02/2022
07/06/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1