It was reported to boston scientific corporation that a resolution clip device was used in the rectum during a colonoscopy procedure performed on (b)(6) 2022.During the procedure, the clip was able to grasp and lock onto tissue; however, the clip was unable to release from the catheter to deploy.The procedure was completed with another resolution clip device.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
|
Block h6: medical device code a15 captures the reportable event of clip unable to release from the catheter.Block h10: investigation results the returned resolution clip device was analyzed, and a visual evaluation noted that the device was returned with the clip assembly attached.It was also observed that the clip had no evidence of activation performed.Microscopic examination was performed and it was found that the clip was deformed since the capsule had hit marks.The clip assembly was flared into the bushing and the locking tabs of the capsule are still closed.Additionally, the control wire was detached from the clip assembly and the bushing had hit marks.Dimensional analysis was performed on the bushing outer diameter, and it was found to be within specification.A dimensional analysis was also performed between the hooks of the bushing and both sides were confirmed to be out of specification.No other problems with the device were noted.The reported event of clip unable to release from the catheter was confirmed.Investigation found that the hooks of the bushing were out of specification, the capsule was damaged, and the device had evidence of flaring of the capsule into the bushing, which can be traced to the manufacturing process as a contributing factor and could have contributed to the clip deployment problem.It is possible that during the manufacturing process, the clip was not assembled correctly, causing the twist, the hit marks on the capsule, and the bushing being out of specification.Regarding the hit marks found on the bushing, this is likely due to the interaction between the yoke and the capsule, in order to deploy the clip caused by the flare of the clip into the bushing.Taking all available information into consideration, the most probable root cause of this event is manufacturing deficiency, as the reported event was traced to the manufacturing process.An investigation to address this problem is in progress.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
|
It was reported to boston scientific corporation that a resolution clip device was used in the rectum during a colonoscopy procedure performed on (b)(6) 2022.During the procedure, the clip was able to grasp and lock onto tissue; however, the clip was unable to release from the catheter to deploy.The procedure was completed with another resolution clip device.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
|