• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. EPIDURAL CATHETERIZATION KIT: 19 GA X 35; ANESTHESIA CONDUCTION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. EPIDURAL CATHETERIZATION KIT: 19 GA X 35; ANESTHESIA CONDUCTION CATHETER Back to Search Results
Model Number IPN923994
Device Problem No Flow (2991)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/22/2022
Event Type  malfunction  
Manufacturer Narrative
Qn#: (b)(4).The device has not been returned for investigation.
 
Event Description
We had two epidurals tonight that the catheters would not flush.We switched to 3 different caps.One epidural i cut the catheter with sterile scissors before the second window on the catheter and then it was able to flush.The other epidural catheter would not flush at all.It was removed and the patient declined a second epidural and then delivered.
 
Event Description
We had two epidurals tonight that the catheters would not flush.We switched to 3 different caps.One epidural i cut the catheter with sterile scissors before the second window on the catheter and then it was able to flush.The other epidural catheter would not flush at all.It was removed and the patient declined a second epidural and then delivered.
 
Manufacturer Narrative
(b)(4).A device history record review was performed on the epidural catheter with no relevant findings.A corrective action is not required at this time as a potential root cause could not be determined based upon the information provided and without a sample.Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed on the epidural catheter with no evidence to indicate a manufacturing related issue.Therefore, the potential cause of this complaint could not be determined based upon the information provided and without a sample.No further action is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIDURAL CATHETERIZATION KIT: 19 GA X 35
Type of Device
ANESTHESIA CONDUCTION CATHETER
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
312 commerce place
asheboro NC 27203
Manufacturer Contact
jasmine brown
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key14305694
MDR Text Key291604243
Report Number1036844-2022-00029
Device Sequence Number1
Product Code BSO
UDI-Device Identifier10801902208986
UDI-Public10801902208986
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K140110
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2023
Device Model NumberIPN923994
Device Catalogue NumberASK-05503-UPM1
Device Lot Number23F22C0037
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/22/2202
Initial Date FDA Received05/06/2022
Supplement Dates Manufacturer Received06/07/2022
Supplement Dates FDA Received06/14/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/30/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
Patient SexFemale
-
-