• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS GOLDWAY (SHENZHEN) INDUSTRIAL INC EFFICIA DFM100 DEFIBRILLATOR/MONITOR; XL+DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS GOLDWAY (SHENZHEN) INDUSTRIAL INC EFFICIA DFM100 DEFIBRILLATOR/MONITOR; XL+DEFIBRILLATOR Back to Search Results
Model Number 866199
Device Problems Image Display Error/Artifact (1304); Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/13/2022
Event Type  malfunction  
Manufacturer Narrative
Reporting address state: (b)(6).Reporting address postal: (b)(6).Reporting institution phone #: (b)(6).Reporter phone: (b)(6).A follow up report will be submitted upon completion of the investigation.
 
Event Description
It was reported to philips that the display is not working.There was no patient involvement.The efficia dfm100 defibrillator, model# 866199, is substantially similar to the heartstart xl+ defibrillator (model # 861290) and will be reported in the united states under device model # 861290.
 
Manufacturer Narrative
Fse evaluated the device during their initial visit and confirmed the customer complaint that the device had malfunctioned.Fse diagnosed the device and determined processor and power printed circuit board (pcb) were defective.Fse initially replaced defective power pcb and processor however the device malfunction was not resolved.Fse reinstalled original power pcb and processor.Fse secondary diagnosis was input/output (i/o) pcb was the source of the failure.Next the fse replaced the i/o pcb, however the unit was still malfunctioning.Final diagnosis of the fse was that the system on module (som) pcb was defective.Finally replacing the som pcb allowed the unit passed testing and the device was returned to use.Parts were not returned for failure investigation (fi).Trends for this type of issue will continue to be monitored to determine if additional actions are required.The complaint will be processed for closure.If additional information becomes available at a later date, the complaint will be reopened, and a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EFFICIA DFM100 DEFIBRILLATOR/MONITOR
Type of Device
XL+DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS GOLDWAY (SHENZHEN) INDUSTRIAL INC
no.2 keji north 3rd road
nanshan district
shenzhen
CH 
Manufacturer (Section G)
PHILIPS GOLDWAY (SHENZHEN) INDUSTRIAL INC
no.2 keji north 3rd road
nanshan district
shenzhen
CH  
Manufacturer Contact
feng she
no.2 keji north 3rd road
nanshan district
shenzhen 
CH  
7552698099
MDR Report Key14313479
MDR Text Key291199632
Report Number3030677-2022-02437
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeSF
PMA/PMN Number
K110825
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number866199
Device Catalogue Number866199
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/13/2022
Initial Date FDA Received05/07/2022
Supplement Dates Manufacturer Received04/13/2022
02/20/2023
Supplement Dates FDA Received06/22/2022
04/19/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/29/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-