Model Number 0998-00-0800-53
Device Problem Circuit Failure (1089)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/11/2022
Event Type  malfunction  
Manufacturer Narrative
Additional information is being requested in regards to whether or not the customer has requested for getinge to service the iabp unit involved.A supplemental report will be submitted when additional information is provided.
 
Event Description
It was reported that the cardiosave intra-aortic balloon pump (iabp) has a possible circuit board issue.It is unknown under which circumstances the event occurred or if a patient was involved; however there was no adverse event reported.
 
Event Description
N/a.
 
Manufacturer Narrative
Updated fields: b4,d9,e1(site country),g3,g6,h2,h3,h4,h6(type of investigation, investigation findings, component codes, investigation conclusions), h10.It was reported that the cardiosave intra-aortic balloon pump (iabp) unit has a possible circuit board issue.It is unknown under which circumstances the event occurred or if a patient was involved; however there was no adverse event reported.A getinge field service engineer (fse) stated that this is duplicate so.Refer to so#44018144 for repair and sign off.Replaced video generator board as per service manual and ensured software compatibility.Unit ran without issue for 30mins.Calibration, safety, and functionality checks passed factory specifcations.Returned to customer upon completion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key14314475
MDR Text Key291296345
Report Number2249723-2022-01062
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/11/2022
Initial Date FDA Received05/07/2022
Supplement Dates Manufacturer Received04/03/2024
Supplement Dates FDA Received04/04/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/30/2015
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-