Updated fields: b4,d9,e1(site country),g3,g6,h2,h3,h4,h6(type of investigation, investigation findings, component codes, investigation conclusions), h10.It was reported that the cardiosave intra-aortic balloon pump (iabp) unit has a possible circuit board issue.It is unknown under which circumstances the event occurred or if a patient was involved; however there was no adverse event reported.A getinge field service engineer (fse) stated that this is duplicate so.Refer to so#44018144 for repair and sign off.Replaced video generator board as per service manual and ensured software compatibility.Unit ran without issue for 30mins.Calibration, safety, and functionality checks passed factory specifcations.Returned to customer upon completion.
|