Model Number CLV-180
Device Problem Component Missing (2306)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
This event was discovered while inspecting a returned loaner device.During this device inspection, it was also noted that the lamp life meter was reading at below 50 hours with an output measuring within range.The scope socket, igniter cable, fan and air pressure were all up to date and functioning properly.If additional information becomes available following the device evaluation, a supplemental report will be filed.
 
Event Description
During receipt inspection of a returned loaner evis exera ii xenon light source, a small screw was noted missing on the back of the main printed circuit board.This event had no patient contact.
 
Manufacturer Narrative
Correction to g3 of the initial medwatch.The aware date should be 11-feb-2022.Investigation activities have been opened to manage the actions related to this report and any required mdr reporting.This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Please see the updates in sections h4, h6, and h10.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 13 years since the subject device was manufactured.Based on the results of the investigation, the investigation confirmed the main board back side was missing a screw.The specific root cause of the missing screw on the main board back side could not be determined at this time.The lens base was cracked and the rear foot deformed; however, these defects are not considered severe enough to cause a potential adverse event.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14337299
MDR Text Key291215719
Report Number8010047-2022-07885
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170215544
UDI-Public04953170215544
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 07/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/11/2022
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/18/2022
Initial Date FDA Received05/09/2022
Supplement Dates Manufacturer Received06/20/2022
Supplement Dates FDA Received07/12/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/03/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-