Device was returned for evaluation.Per the manufacturer's evaluation findings: unable to confirm a defect with the light fibers.Scrapes and debris in the working channel.The lightguides are intact, however, there are some visible voids in the epoxy and some of the fibers appear to protrude from the distal face; could not confirm if the fibers are loose or broken off.Although the lightguides appear to be raised from the distal face, there were no signs of loose fibers large enough to have been visible inside the patient's bladder.Additional findings: chipped distal ceramic sleeve.External puncture in the angle cover 50mm from distal tip, leaks, and debris on the light taper threads.
|