BAXTER HEALTHCARE CORPORATION; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
|
Back to Search Results |
|
Catalog Number ASKU |
Device Problem
Fitting Problem (2183)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/13/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was not returned and the lot number is unknown; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that there was a connection issue between the titanium adapter and the patient adapter of the peritoneal dialysis (pd) transfer set.The connection issue was further described as, ¿unable to connect tightly¿.This was observed during use of the device for pd therapy.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Search Alerts/Recalls
|
|
|