It was reported on (b)(6) 2022 that the patient received electric shocks from the mx450 device.When the device put the plug in the mains, it caused a spark.This had happened to several users.The device rack also receives electric shocks when connected to the mains.The device was in normal use (except the power cord is probably replaced with ragged power cord not delivered with the medical equipment per biomed).The remote service engineer (rse) was consulted on the case and was able to confirm that the device was pulled from usage and no injury or treatment was had from the issue.The rse spoke with the biomed onsite and was able to determine that the wrong cable for the device was being used when the issue occurred.The mx450 was sent to bench repair and received on may 11th, 2022 to confirm the issue of the cable being the cause of the shock.After arriving to bench, the used cable was confirmed to be an alternative power cable (not the one provided by philips) with a broken earth conductor, thus it didn¿t pass electrical safety tests when being plugged into a wall connector and exhibited sparks as stated.The device was working as expected after the bench repair with no internal connections or groundings being damaged.The not original non philips power cord was damaged and replaced with the original power cord.The issues were found with the mx450 and was deemed fully functional.
|