Model Number 560BCS1 |
Device Problem
Electrical /Electronic Property Problem (1198)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/27/2022 |
Event Type
malfunction
|
Event Description
|
During preventive maintenance by a service technician this bio-console 560 instrument required service/repair due to a fail during system controller replacement.This was detected during service so there was no patient involvement, so no adverse effect occurred.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Device evaluation summary: during preventive maintenance by a service technician for technical service update (tsu), the service technician observed an error 69 when the instrument started up and displayed a ups malfunction.The service technician reported that the dc power supply cable had been burnt.During service, technician did not disconnect batteries before system controller board replacement and caused a short circuit on the pcb.The issue was resolved by replacing the base unit 560 cable assy ps and hw, screw, 8, ph, pl, ss, 1.0.The assy system controller module -006 was replaced per tsu.Preventive maintenance was completed per specifications.Additional information: medtronic received additional information that the issue occurred during an update related to tsu mod0206.The service technician did not disconnect the batteries before system controller board replacement, causing a short circuit on the pcb.After assembly, the bio-console did not work at all.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
(eval code conclusion/fdc/annex d): this field has been updated to d02.(usage of device): this field has been updated to reuse.Correction additional codes (imf): this field has been updated to f27.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|