• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC T.W. POWER SUPPLY; UNIT, CAUTERY, THERMAL, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC T.W. POWER SUPPLY; UNIT, CAUTERY, THERMAL, AC-POWERED Back to Search Results
Model Number T.W. POWER SUPPLY
Device Problem Failure to Deliver Energy (1211)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/18/2022
Event Type  malfunction  
Manufacturer Narrative
Trackwise id: (b)(4).
 
Event Description
The hospital reported that during an endoscopic vein harvesting procedure the t.W.Power supply was making an audible tone during use and sealing wouldn't occur when the trigger was engaged on the handpiece.The power supply was swapped out and the case was completed with the same disposable in use.The case was completed successfully.No patient effects occurred.
 
Manufacturer Narrative
Trackwise id # (b)(4).The reported device is an oem device, therefore, a lot history review was not applicable.The product is not returning.A lot number was not provided and the specific product lot number cannot be identified from a ship history, therefore it is impossible to obtain a certificate of conformance.The device was not returned to maquet cardiac surgery for investigation, therefore no evaluation could be performed.It is not possible to confirm the reported complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T.W. POWER SUPPLY
Type of Device
UNIT, CAUTERY, THERMAL, AC-POWERED
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key14346698
MDR Text Key291307808
Report Number2242352-2022-00393
Device Sequence Number1
Product Code HQO
UDI-Device Identifier00607567700826
UDI-Public00607567700826
Combination Product (y/n)N
PMA/PMN Number
K052274
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 07/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberT.W. POWER SUPPLY
Device Catalogue NumberVH-3010
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/18/2022
Initial Date FDA Received05/10/2022
Supplement Dates Manufacturer Received07/22/2022
Supplement Dates FDA Received07/22/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-