• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED CELLEX PHOTOPHERESIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED CELLEX PHOTOPHERESIS SYSTEM Back to Search Results
Catalog Number CLXECP
Device Problem Obstruction of Flow (2423)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/29/2021
Event Type  malfunction  
Event Description
The customer contacted mallinckrodt to report they experienced clots observed in the return line with their cellex photopheresis kit ("kit") during an extracorporeal photopheresis (ecp) treatment.The customer reported they observed blood clotting in the return line.The ecp treatment was aborted and residual blood within the kit was not returned to the patient.The customer reported the patient was in stable condition.The customer did not return the product for investigation.
 
Manufacturer Narrative
The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction clot observed in the return line.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review for kit lot j150 was conducted.There were no non-conformances associated with this lot.This lot met all release requirements.A review of kit lot j150 shows no trends.Trends were reviewed for complaint category, clot observed.No trends were detected for this complaint category.The assessment is based on the information available at the time of investigation.No product was returned for investigation; therefore, it could not be determined if the product met specification based solely on the information provided by the customer.Section 2-9 of the cellex operators manual (1470096 rev c) for use with software 5.1 on anticoagulant states "caution: individual patients may require a heparin dosage that varies from the recommended dose to prevent post-treatment bleeding or clotting during a treatment.The physician should review the patient's medical condition, medications and platelet count at the time of treatment and use clinical judgement to establish the optimal heparin dosage for each patient." the root cause for the clotting observed could not be determined based on the available information.No further action is required at this time.This investigation is now complete.Comp- (b)(4) p.T.(b)(6) 2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELLEX PHOTOPHERESIS SYSTEM
Type of Device
CELLEX PHOTOPHERESIS SYSTEM
Manufacturer (Section D)
MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED
blanchardstown, dublin D15 T X2V
EI  D15 TX2V
Manufacturer (Section G)
MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED
college business & tech park
cruiserath road
blanchardstown, dublin D15 T X2V
EI   D15 TX2V
Manufacturer Contact
megan vernak
shelbourne building
53 frontage rd suite 300
hampton, NJ 08827
MDR Report Key14350134
MDR Text Key299872818
Report Number3013428851-2022-00036
Device Sequence Number1
Product Code LNR
Combination Product (y/n)Y
Reporter Country CodeCO
PMA/PMN Number
P860003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 05/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/01/2022
Device Catalogue NumberCLXECP
Device Lot NumberJ150
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/06/2022
Initial Date FDA Received05/10/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/04/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexFemale
-
-