• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. DREAMSTATION AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. DREAMSTATION AUTO CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number DSX500T11C
Device Problems Contamination (1120); Material Erosion (1214); Nonstandard Device (1420); Particulates (1451)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/10/2021
Event Type  malfunction  
Event Description
The manufacturer received information alleging seeing black particels in the hose related to a cpap device's sound abatement foam.There was no report of patient harm or injury.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Manufacturer Narrative
The manufacturer was previously received information alleging particles in tubing/chamber, difficulty breathing/short of breath, headaches that have been very bad waking her up out of sleep and its concerning related to a cpap device's sound abatement foam.There was no report of patient harm or injury.The device was returned to the manufacturer's product investigation laboratory for investigation.An internal visual inspection of the device was completed by the manufacturer.The manufacturer found evidence of white talc/dried white powder.The manufacturer found no evidence of sound abatement foam degradation/breakdown.The device's event logs were downloaded and reviewed.The manufacturer found no error logged.The manufacturer concludes there was evidence of white talc/dried white powder contamination on the device.The manufacturer confirmed there was no evidence of sound abatement foam degradation/breakdown.Section d9, g3 and h6 has been updated/corrected in this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION AUTO CPAP
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
2673970028
MDR Report Key14365770
MDR Text Key299887539
Report Number2518422-2022-18417
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date 03/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberDSX500T11C
Device Catalogue NumberDSX500T11C
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 06/10/2021
Initial Date FDA Received05/11/2022
Supplement Dates Manufacturer Received03/27/2023
Supplement Dates FDA Received03/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/20/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberRES 88058
Patient Sequence Number1
-
-