• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE Back to Search Results
Model Number GIF-HQ190
Device Problems Device Reprocessing Problem (1091); Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/14/2022
Event Type  malfunction  
Manufacturer Narrative
The subject device was received and evaluated.Device inspection , and as requested by the customer, service repair performed forceps passage check and found tear marks on walls of the channel.In addition, the distal end plastic cover found with dents and scratches.The a-rubber (bending section) c-cover was noted with crack including the (i/t) insertion tube.Investigation is ongoing.This report will be supplemented accordingly following investigation.
 
Event Description
It was reported during reprocessing , metal debris was expelled from the device.According to the customer, the unit passed leak test however, would like an inspection of the device to ensure no internal damage exists.There was no patient harm or injury reported.No user injury reported.
 
Manufacturer Narrative
Correction to e3, the reporters title was biomedical engineer.This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause could not be determined.However, that it is likely that a part of an accessory was broken and exited from the biopsy channel due to inappropriate accessory handling by the user.The following is included in the instructions for use (ifu) and may have helped prevent the event: "inspection of the instrument channel" "if insertion or withdrawal of endotherapy accessories is difficult, straighten the bending section as much as possible without losing the endoscopic image.Inserting or withdrawing endotherapy accessories with excessive force may damage the instrument channel or endotherapy accessories and could cause some parts to fall off and/or cause patient injury." "if the distal end of an endotherapy accessory is not visible in the endoscopic image, do not open the distal end or extend the needle of the endotherapy accessory.This could cause patient injury, bleeding, perforation, and/or equipment damage." "for more information on combining the endoscope with particular endotherapy accessories, refer to combination equipment and the instruction manuals for the accessories.Also, refer to their respective instruction manuals for operating the accessories." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE
Type of Device
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14366388
MDR Text Key291512523
Report Number8010047-2022-08093
Device Sequence Number1
Product Code FDS
UDI-Device Identifier04953170305276
UDI-Public04953170305276
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131780
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 07/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGIF-HQ190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/14/2022
Initial Date FDA Received05/11/2022
Supplement Dates Manufacturer Received06/14/2022
Supplement Dates FDA Received07/11/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-