• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS DMC GMBH DIGITALDIAGNOST DUAL DETECTOR; SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS DMC GMBH DIGITALDIAGNOST DUAL DETECTOR; SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER) Back to Search Results
Model Number 712022
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that the patient support cannot be fixed.No injury occurred.
 
Manufacturer Narrative
Ref.Id: (b)(4) the investigation is still ongoing on this event.When the investigation is completed a follow-up report will be sent to the fda.
 
Manufacturer Narrative
Ref.Id: (b)(4).The digitaldiagnost is a direct digital radiography system with flat detector technology based on modular components to allow customization for all radiographic applications and workload requirements.For anatomies that are larger than the detector size like scoliosis, it is possible to make a series of exposures covering the whole anatomy (for example full spine or full legs).These individual images can be "stitched" together via the software program.For proper patient positioning and support, the patient can be placed on the so called patient support for stitching or stitching stand, a platform with two handles that the patient can hold on to during the examination.For the ease of use during transportation, the patient support for stitching has wheels and a folding footboard.The footboard is connected to the frame via two hinges and a brake cylinder.It has to be folded up and fixed by a hook for transportation, e.G.From one room to another.Philips field service engineer visited the customer site.The mounting of brake cylinder was detached and cannot be fixed on the frame, the intended position has been deformed.Hence field service engineer reinstalled the screws tightly and checked the system.System works as specified.Risk estimation revealed no unacceptable risk.The issue is further monitored and trended.The event was originally reported to the authorities as not enough information was available to make a definite reportability decision.Since that time, additional information was received which revealed that the event does not meet the criteria for reporting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIGITALDIAGNOST DUAL DETECTOR
Type of Device
SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS DMC GMBH
roentgenstrasse 24
hamburg 22335
GM  22335
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS DMC GMBH
roentgenstrasse 24
hamburg 22335
GM   22335
Manufacturer Contact
dusty leppert
222 jacobs st
cambridge, MA 02141
6172455900
MDR Report Key14372716
MDR Text Key291492794
Report Number3003768251-2022-00005
Device Sequence Number1
Product Code MQB
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K090625
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number712022
Device Catalogue Number712022
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/14/2022
Initial Date FDA Received05/12/2022
Supplement Dates Manufacturer Received04/14/2022
Supplement Dates FDA Received09/21/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-