Model Number CV-180
Device Problem Key or Button Unresponsive/not Working (4063)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/13/2022
Event Type  malfunction  
Manufacturer Narrative
The device was returned, and an initial evaluation was conducted by olympus; however, investigation is ongoing.During the initial evaluation, the user's report was confirmed.A faulty power supply switch was discovered and causing the power button to not stay on.Additionally, the front panel standoffs were noted damaged and the intermittent eject button on the printed circuit board was compromised.If additional information becomes available following the device evaluation, a supplemental report will be filed.
 
Event Description
The customer originally returned his olympus evis exera ii video system center due to a power button issue.According to the initial reporter, when attempting to power on the device, the power button would not stay pushed in.Reportedly, this did not take place during a procedure.No patient involvement for this event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 11 years since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, the root cause of the faulty power supply switch could not be identified.However, it¿s likely the cause is related to an accidental failure or a failure caused by handling (excessive stress to the switch).The reported event might have been prevented by following the instructions for use below: [important information ¿ please read before use].Do not apply excessive force to this video system center and/or other instruments connected.Otherwise, damage and/or malfunction can occur.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II VIDEO SYSTEM CENTER
Type of Device
VIDEO SYSTEM CENTER
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14373163
MDR Text Key291494562
Report Number8010047-2022-08114
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170215513
UDI-Public04953170215513
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCV-180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/28/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/21/2022
Initial Date FDA Received05/12/2022
Supplement Dates Manufacturer Received06/30/2022
Supplement Dates FDA Received07/26/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/26/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-