Model Number CV-180 |
Device Problem
Key or Button Unresponsive/not Working (4063)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/13/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned, and an initial evaluation was conducted by olympus; however, investigation is ongoing.During the initial evaluation, the user's report was confirmed.A faulty power supply switch was discovered and causing the power button to not stay on.Additionally, the front panel standoffs were noted damaged and the intermittent eject button on the printed circuit board was compromised.If additional information becomes available following the device evaluation, a supplemental report will be filed.
|
|
Event Description
|
The customer originally returned his olympus evis exera ii video system center due to a power button issue.According to the initial reporter, when attempting to power on the device, the power button would not stay pushed in.Reportedly, this did not take place during a procedure.No patient involvement for this event.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 11 years since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, the root cause of the faulty power supply switch could not be identified.However, it¿s likely the cause is related to an accidental failure or a failure caused by handling (excessive stress to the switch).The reported event might have been prevented by following the instructions for use below: [important information ¿ please read before use].Do not apply excessive force to this video system center and/or other instruments connected.Otherwise, damage and/or malfunction can occur.Olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|
|