• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, LLC RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, LLC RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number V60
Device Problem Unable to Obtain Readings (1516)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/02/2022
Event Type  malfunction  
Event Description
It was reported to philips the tidal volume was not registering.This event was reported to have occurred during patient use.The patient was swapped to a different ventilator.Further information has been requested.If additional information becomes available at a later date, a supplemental report will be submitted.
 
Manufacturer Narrative
The philips field service engineer (fse) confirmed and duplicated the issue.The device was reading the air flow at 240 lpm at all times.Following the evaluation of the device, the fse replaced the gas delivery system (gds) to resolve the problem.The unit was then functionally tested and successfully passed specified tests.No other abnormality was observed.
 
Manufacturer Narrative
The gas delivery subsystem (gds) was returned to philips for failure investigation (fi).The customer complaint was verified.The root cause was failure of the airflow sensor, caused by u1 drifting out of calibration.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
RESPIRONICS CALIFORNIA, LLC
2271 cosmos court
carlsbad CA 92011
Manufacturer Contact
kimberly shelly
2271 cosmos court
carlsbad, CA 92011
7609187300
MDR Report Key14380738
MDR Text Key292368049
Report Number2031642-2022-01282
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public00884838020054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/02/2022
Initial Date FDA Received05/12/2022
Supplement Dates Manufacturer Received06/01/2022
11/22/2022
Supplement Dates FDA Received06/23/2022
11/30/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/25/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-