RESPIRONICS CALIFORNIA, LLC RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Intermittent Loss of Power (4016)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/28/2022 |
Event Type
malfunction
|
Event Description
|
It was reported to philips the unit powers on by itself intermittently.The device did not have patient involvement at the time the issue was discovered, there was no patient or user harm reported.Philips remote service engineer discussed the signal path for the power switch per the service manual and provided the part numbers for the power switch overlay and for the pm pcba.Further information has been requested.If additional information becomes available at a later date, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
The customer called back to report the replacement of the front panel overlay and power management (pm) board did not resolve the issue.The remote service engineer (rse) informed the customer that the issue could potentially be the user interface (ui) printed circuit board assembly (pcba) or the central processing unit (cpu).The customer has requested part numbers of all potential parts.Multiple attempts were made to follow-up with the customer to obtain repair information; however, there was no response.
|
|
Search Alerts/Recalls
|
|
|