• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. HIGH FLOW INSUFFLATION UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-4
Device Problem Failure to Auto Stop (2938)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/25/2022
Event Type  malfunction  
Manufacturer Narrative
As stated this event was discovered during a returned loaner device inspection.Upon further evaluation, it was confirmed that the auto shut down was failing due to a faulty printed circuit board.If additional information becomes available following the device evaluation, a supplemental report will be filed.
 
Event Description
During the evaluation of a returned loaner olympus high flow insufflation unit, it was discovered that the unit's auto shut down function had failed.This event did not have any patient involvement.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 8 years since the subject device was manufactured.Based on the results of the investigation, the phenomenon (the unit auto shut down function failed) occurred due to failure of the printed circuit board.The printed circuit board malfunctioned within less than a year since the previous repair but its cause could not be identified.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14407775
MDR Text Key291724329
Report Number8010047-2022-08312
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170324147
UDI-Public04953170324147
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 06/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/29/2022
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/25/2022
Initial Date FDA Received05/16/2022
Supplement Dates Manufacturer Received06/02/2022
Supplement Dates FDA Received06/29/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/23/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-