• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065752452
Device Problems Insufficient Flow or Under Infusion (2182); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/28/2022
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A physician reported infusion was unstable and connecting part of an infusion tube came off during a procedure.The procedure was completed after the product was replaced.There was no patient harm.
 
Manufacturer Narrative
Additional information provided in d.9., h.3., h.6., and h.10.A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The customer reported that the infusion auto infusion valve (aiv) filter was broken in their pak.One used pak was visually inspected.The housing of the auto stopcock fell apart.The top and base auto stopcock housing appeared to have been insufficiently welded.The root cause of the customer's complaint was determined to be related to equipment not maintaining consistent control of the weld process.Contributing factors identified were training issues during production when performing the integrity squeeze check on the component to verify a good weld and when performing maintenance on the equipment.Action was taken to determine root cause and to take appropriate corrective actions for complaints of a similar nature.Quality assurance identified the likely contributors to be related to equipment end of life and will be replaced with a new system.Procedural enhancements were also implemented for better control of process during quality checks and for better training of personnel during production and equipment maintenance.Complaints are reviewed and monitored at regular intervals for adverse trends.No further action has been identified for this reported event.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14427445
MDR Text Key291943262
Report Number1644019-2022-00377
Device Sequence Number1
Product Code LRO
UDI-Device Identifier00380657524525
UDI-Public00380657524525
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2023
Device Catalogue Number8065752452
Device Lot Number2487729H
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/08/2022
Initial Date FDA Received05/18/2022
Supplement Dates Manufacturer Received08/04/2022
Supplement Dates FDA Received08/23/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CONSTELLATION VISION SYSTEM
-
-