BAXTER HEALTHCARE CORPORATION MINICAP TRANSFER SET; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
|
Back to Search Results |
|
Catalog Number 5C4482 |
Device Problem
Fitting Problem (2183)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/31/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that a minicap extended life pd transfer set had a connection issue between the patient connector and the adapter.This occurred ¿before treatment¿ for peritoneal dialysis therapy.There was no allegation against the adapter.The titanium adapter remained in use.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.The photograph was visually inspected and showed the transfer set inside an opened pouch with tip protectors in place.The blue pull ring cap is in place over the patient connector (luer adapter), the patient connector could not be seen for a detailed visual inspection.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|