Model Number M00522610 |
Device Problem
Activation, Positioning or Separation Problem (2906)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/22/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Initial reporter address: (b)(6).(b)(4).The device has not been received for analysis.Upon receipt and completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
|
Event Description
|
It was reported to boston scientific corporation that a resolution clip device was used in the colon during an endoscopic mucosal resection (emr) procedure performed on (b)(6) 2022.During the procedure, the clip could not detach from the catheter after it was deployed.The procedure was completed with another resolution clip device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
|
Manufacturer Narrative
|
Block e1: initial reporter address: (b)(6).Block h6: medical device code a15 captures the reportable event of clip could not detach from the catheter.Block h10: investigation results: the returned resolution clip device was analyzed, and a visual evaluation noted that the device was returned with the clip assembly attached.The device was returned without the over-sheath.Microscopic examination was performed and it was found that the clip arms were misaligned.The clip had no activation performed.Functional evaluation was performed and the clip assembly can open its arms within specification under the tortuous path.Additionally, the clip assembly was able to perform the first activation and could be released from the catheter without problem.No other problems with the device were noted.The reported event of clip could not detach from catheter was not confirmed.Investigation found that the clip arms were misaligned, this is likely due to operational factors such as the amount of tissue grasped during the procedure.The returned device review showed no evidence of either the alleged(s) or any defect which could have contributed to the event.Therefore, the most probable root cause for this complaint is no problem detected.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
|
|
Event Description
|
It was reported to boston scientific corporation that a resolution clip device was used in the colon during an endoscopic mucosal resection (emr) procedure performed on (b)(6) 2022.During the procedure, the clip could not detach from the catheter after it was deployed.The procedure was completed with another resolution clip device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
|
Search Alerts/Recalls
|