• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. VISERA ELITE XENON LIGHT SOURCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. VISERA ELITE XENON LIGHT SOURCE Back to Search Results
Model Number CLV-S190
Device Problems Electrical /Electronic Property Problem (1198); Output below Specifications (3004)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/21/2022
Event Type  malfunction  
Manufacturer Narrative
The referenced device was returned to the repair center for evaluation.The customer¿s reported issue was confirmed; the e200 error was attributed to a defective/faulty filter turret and the front panel indicator flashed.The root cause of the faulty turret cannot be confirmed.A device has not been repaired in the past year.The legal manufacturer performed a review of the device history records for the concerned device and no abnormalities were found.In general, the customer is required to check the function of all devices used prior to a procedure and as a preventative measure, a suitable replacement device must be provided during an application.Olympus will continue to monitor complaints for this device.
 
Event Description
Olympus (osh) was informed by the user facility via service inquiry that the customer visera elite xenon light source displays a reported ¿e200, front panel indicators flash and the lamp is dark¿.The event was found during reprocessing.There was no delay as the scheduled laryngoscopy was completed using another similar light source.No death, injury or harm was reported to olympus.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and to correct information provided on the initial report.The following sections were corrected: g2, specify other as ¿china¿.It has been over 8 years since the subject device was manufactured.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA ELITE XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14459114
MDR Text Key300330516
Report Number8010047-2022-08591
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170301995
UDI-Public04953170301995
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-S190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/23/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/21/2022
Initial Date FDA Received05/20/2022
Supplement Dates Manufacturer Received05/23/2022
Supplement Dates FDA Received06/20/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/10/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-