During a new patient implant, the surgeon opted to not use the original lead that was going to be implanted as he believed it was bent.The livanova representative that was present noted that the surgeon was working on the negative and positive electrodes for about 20 minutes.While visually inspecting the lead, the rep noticed that the negative electrode is exposed in the coil.The suspect lead was received but product analysis is still underway.No other relevant information has been received to date.
|
Product analysis was completed on the returned lead.The allegation of ¿mechanical problem, damaged/kinked coil¿ was not confirmed.During the visual analysis kinks on coils were not observed, however five atypical marks on connector pin, were observed.Also, a scratch on connector ring leading to a cut off portion of first large o-ring was observed.Based on the appearance of the returned lead it is believed that the above-mentioned observations were most likely caused during manipulation of the lead at surgery.Continuity checks of the returned lead portion was performed during the functional analysis, and no discontinuities were identified.Other than the noted above conditions, the condition of the returned lead portion is consistent with conditions that typically exist following an implant/explant procedure.
|