• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD R3 42MM ID US COCR LNR 54MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/CERAMIC/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD R3 42MM ID US COCR LNR 54MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/CERAMIC/METAL Back to Search Results
Catalog Number 71341154
Device Problem Biocompatibility (2886)
Patient Problems Pain (1994); Synovitis (2094); Metal Related Pathology (4530)
Event Date 04/02/2021
Event Type  Injury  
Event Description
It was reported that, after an r3-tha construct had been implanted on the patient¿s right hip on (b)(6) 2009, the patient experienced metallosis and pain.This adverse event was treated with a revision surgery on (b)(6) 2021.During this procedure, all components were explanted and replaced with a competitors¿ tha system.Intraoperatively, inflammatory synovial changes were noticed with a small gross macroscopic area with tissue staining.The trunnion showed mild signs of corrosion.The patient was transferred in a stable condition to the recovery room.
 
Manufacturer Narrative
Internal complaint reference (b)(4).
 
Manufacturer Narrative
H3, h6: it was reported that a right hip revision surgery was performed due to metallosis and pain.Intraoperatively, inflammatory synovial changes and trunnion mild signs of corrosion were noticed.As of today, the implanted devices, all of which were used in treatment have not been returned for evaluation.A review of the historical complaints data for the metal liner was performed using batch numbers, part numbers and the reported failure modes to evaluate patterns of repeated failures or defects.Similar complaints were identified for the metal liner, however, as the device is no longer sold, no action is to be taken.In the absence of the actual device, the production records were reviewed for the devices reportedly involved in this incident.All released devices involved met manufacturing specifications upon release for distribution.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.No further actions are required at this time.A review of historic escalation actions related to the product and similar complaint events was performed.Following the review, prior applicable escalation actions were identified and confirmed to reduce associated risks as far as possible.No further escalation actions are required.The available medical documents were reviewed.The clinical information provided, of tissue staining and corrosion on the trunnion may be consistent with a reaction to metal debris.However, the source and the root cause cannot be determined with the available documentation.It cannot be concluded that the reported clinical findings are associated with the implant failure.Based on the available information we can confirm the reported complaint, however without further information our investigation remains inconclusive, and a definitive root cause cannot be determined.Specific factors known to contribute to the alleged fault are excessive physical activity levels, unreasonable stress on replacement system, excessive patient weight, trauma to the joint replacement, loosening of components may increase production of wear particles and accelerate damage to the bone.Should the device or additional information be received, the complaint will be reopened.Based on this investigation, the need for corrective and preventative actions is not indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 42MM ID US COCR LNR 54MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/CERAMIC/METAL
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK   CV31 3HL
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14467826
MDR Text Key292349988
Report Number3005975929-2022-00322
Device Sequence Number1
Product Code MRA
UDI-Device Identifier00885556022351
UDI-Public885556022351
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/29/2018
Device Catalogue Number71341154
Device Lot Number08MW20435
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/28/2022
Initial Date FDA Received05/23/2022
Supplement Dates Manufacturer Received09/14/2022
Supplement Dates FDA Received09/15/2022
Date Device Manufactured12/31/2008
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
Patient Age55 YR
Patient SexMale
-
-