Model Number 712027
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The mounting of brake cylinder from the footplate of the stitching stand came off and the footplate struck the technician.
 
Manufacturer Narrative
Ref.Id: (b)(4).The investigation is still ongoing on this event.When the investigation is completed a follow-up report will be sent to the fda.
 
Manufacturer Narrative
Ref.Id: (b)(4).The digitaldiagnost is a direct digital radiography system with flat detector technology based on modular components to allow customization for all radiographic applications and workload requirements.For anatomies that are larger than the detector size like scoliosis, it is possible to make a series of exposures covering the whole anatomy (for example full spine or full legs).These individual images can be "stitched" together via the software program.For proper patient positioning and support, the patient can be placed on the so called patient support for stitching.For the ease of use during transportation, the patient support for stitching has wheels and a folding footboard.The footboard is connected to the frame via two hinges and a brake cylinder.The mounting of the lower part of the brake cylinder was just detached from the base frame, the probable cause could not be determined.The injury to the tech was not confirmed.The field service engineer corrected the issue by refixing the foot board mounting plate with the brake cylinders and checked the system.System works as specified.Risk estimation revealed no unacceptable risk.The issue is further monitored and trended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIGITALDIAGNOST 4 HIGH PERFORMANCE
Type of Device
SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS DMC GMBH
roentgenstrasse 24
hamburg 22335
GM  22335
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS DMC GMBH
roentgenstrasse 24
hamburg 22335
GM   22335
Manufacturer Contact
dusty leppert
222 jacobs st
cambridge, MA 02141
6172455900
MDR Report Key14468729
MDR Text Key292356519
Report Number3003768251-2022-00006
Device Sequence Number1
Product Code MQB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141736
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number712027
Device Catalogue Number712027
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/22/2022
Initial Date FDA Received05/23/2022
Supplement Dates Manufacturer Received04/22/2022
Supplement Dates FDA Received02/03/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-