Customer report of sparking for afinion as100 analyzer while doing a patient test.Sparking was seen between the power cord and the analyser, the customer was unable to specify if spark was from power cord or analyzer.Analyzer was located in a dry location at 71°f (21/22°c), on a flat surface.Power cord was connected to a surge protector.There was no: visible smoke, evidence of smoke, static or electric shock, charing, burnt smell, scorch mark or fire.There was no injury or death as a result of the sparking.There was no delay or impact to patient test / treatment, test was performed on another analyzer.
|
Customer report of sparking for afinion (b)(6) analyzer while doing a patient test.Sparking was seen between the power cord and the analyser, the customer was unable to specify if spark was from power cord or analyzer.Analyzer was located in a dry location at 71°f (21/22°c), on a flat surface.Power cord was connected to a surge protector.There was no: - visible smoke.- evidence of smoke.- static or electric shock.- charing.- burnt smell.- scorch mark.- fire.There was no injury or death as a result of the sparking.There was no delay or impact to patient test / treatment, test was performed on another analyzer.
|
The instrument was to be returned for further investigation.The instrument still has not been received in oslo.Therefore, return testing has not been performed.The investigation determined that the reported event is not related to product results, inadequate instructions for use or inaccurate labels.
|