Model Number CD2357-40Q
Device Problem Migration (4003)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  Injury  
Event Description
Related manufacturer reference number: 2017865-2022-10772 / 2017865-2022-10773.It was reported that the patient presented to the emergency room with phrenic nerve stimulation.Upon review, the right atrial(ra) and right ventricular(rv) lead were dislodged and the implantable cardioverter defibrillator moved due to suspected twiddler's syndrome.Dislodgement was confirmed via x-ray imaging.The rv lead was re-positioned successfully on (b)(6) 2022.The ra lead and the device were explanted and replaced.The patient condition was stable post-procedure.
 
Manufacturer Narrative
It was reported that the device migrated due to twiddlers syndrome.Interrogation of the device revealed the device was above elective replacement indicator (eri) when received.Functional testing was normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORTIFY ASSURA DR ICD, US
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key14476890
MDR Text Key292476998
Report Number2017865-2022-10778
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier05414734508094
UDI-Public05414734508094
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2022
Device Model NumberCD2357-40Q
Device Catalogue NumberCD2357-40Q
Device Lot NumberA000092479
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/16/2022
Initial Date FDA Received05/23/2022
Supplement Dates Manufacturer Received06/28/2023
Supplement Dates FDA Received07/07/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/24/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age48 YR
Patient SexFemale
-
-