Model Number WA00014A
Device Problem Sparking (2595)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
Olympus was informed that during six different therapeutic transurethral resection of the prostate (turp) procedures at an unknown date six hf cables from the same lot-number sparked.No further information was provided but there was no report of an adverse event or patient injury.
 
Manufacturer Narrative
The suspect medical device has not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the user's experience and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.Please note: the report id for the first report was 9610773-2022-00173.
 
Manufacturer Narrative
The suspect medical device was returned to the manufacturer for evaluation/investigation.The investigation found all stranded wires inside the cable to be broken and melted at one of the connectors.This is a known fault pattern.Wear and tear due to frequent use in connection with repeated great bending and/or tensile loads most likely caused the stranded wires inside the cable to break causing voltage flashovers and the reported electric sparking when the hf generator was activated.According to the article¿s lot number, the hf cable was manufactured in december 2020.It is assumed that the hf cable was used longer than the 12 months the cable is designed for.Thus, this event/incident can most likely be attributed to use-related wear and tear in connection with improper handling by the customer and thus to use error.A material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot number of the hf cable without showing any abnormalities.The case will be closed on olympus side with no further actions.However, the reported event/incident will be recorded for trending and surveillance purposes and the user will be informed about the investigation results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-CABLE, BIPOLAR
Type of Device
HF-CABLES
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM  22045
Manufacturer (Section G)
MED CONTACT GMBH
kornbühlstr. 100-102
salmendingen
Manufacturer Contact
daniel wladow
kuehnstrasse 61
hamburg 22045
GM   22045
4940669662
MDR Report Key14485911
MDR Text Key292726281
Report Number9610773-2022-00189
Device Sequence Number1
Product Code FAS
UDI-Device Identifier04042761076449
UDI-Public04042761076449
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K120418
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWA00014A
Device Catalogue NumberWA00014A
Device Lot Number20ZW
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/10/2022
Initial Date FDA Received05/24/2022
Supplement Dates Manufacturer Received06/10/2022
Supplement Dates FDA Received06/14/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexMale
-
-