BOSTON SCIENTIFIC CORPORATION EXPRESS LD ILIAC / BILIARY; STENT, ILIAC Back to Search Results
Model Number 20226
Device Problems Difficult to Remove (1528); Failure to Advance (2524); Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/06/2022
Event Type  Injury  
Event Description
It was reported that stent dislodgement inside the patient and removal difficulties were encountered.The 99% stenosed target lesion was located in the mildly tortuous and severely calcified common iliac.A 9.0x40x75cm express ld stent was advanced for treatment.However, during the procedure, the stent delivery system failed to cross the lesion and then became difficult to remove from the lesion.When attempted to remove the device, the stent dislodged from the balloon.An attempt was made to deploy the dislodged stent with the balloon but it was unsuccessful.The balloon delivery system was removed.A non-bsc stent was used to affix and appose the dislodged express stent to the vessel wall.The procedure was completed with no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRESS LD ILIAC / BILIARY
Type of Device
STENT, ILIAC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key14492825
MDR Text Key292568020
Report Number2134265-2022-05460
Device Sequence Number1
Product Code NIO
UDI-Device Identifier08714729392330
UDI-Public08714729392330
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133110
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 05/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date04/15/2024
Device Model Number20226
Device Catalogue Number20226
Device Lot Number0027153933
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/06/2022
Initial Date FDA Received05/24/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/16/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexFemale
Patient RaceWhite
-
-