• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET MICROFIXATION UNKNOWN MANDIBULAR SCREW

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET MICROFIXATION UNKNOWN MANDIBULAR SCREW Back to Search Results
Model Number N/A
Device Problems Device Dislodged or Dislocated (2923); Migration (4003)
Patient Problem Pain (1994)
Event Type  Injury  
Manufacturer Narrative
Zimmer biomet complaint #: (b)(4).Concomitant medical products: marahar pm-tmj & model catalog #: tmjpm-2685 lot #: 949090a foreign: (b)(6).The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated report: 0001032347-2022-00167.
 
Event Description
It was reported that patient had original surgery to implant custom tmj prostheses approximately two years ago.Subsequently, all of the screws have pulled out and have radiolucencies around them.The surgeon stated that this suggests a systemic bone healing problem.The patient will need a revision, but the surgeon is having her worked up for a bone physiology disorder first as this could be part of her mctd or something different.Attempts have been made and no further information has been provided.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.2019 pre-op panorex, 2019 post-op panorex, april 2022 panorex and ct images were provided and reviewed by a health care professional.Review of the available images identified the following: pre-op imaging demonstrated suspected bilateral tmj arthrosis.2019 post-op imaging demonstrates anatomic alignment of bilateral custom tmj arthroplasties without abnormality.April 2022 imaging demonstrates loosening of the bilateral tmj arthroplasties with lucencies along the bilateral screws and slight retraction of the mandibular implants from the osseous margins.There is no osseous or hardware fracture.Patient anatomy appears unremarkable and unchanged, apart from bone quality appears mildly osteopenic.Part and lot identification are necessary for review of device history records, neither were provided for the screws.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN MANDIBULAR SCREW
Type of Device
SCREW
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
Manufacturer (Section G)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key14497738
MDR Text Key292696138
Report Number0001032347-2022-00168
Device Sequence Number1
Product Code DZL
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberUNKNOWN MANDIBULAR SCREWS
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/02/2022
Initial Date FDA Received05/25/2022
Supplement Dates Manufacturer Received09/13/2022
Supplement Dates FDA Received09/30/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age39 YR
Patient SexFemale
-
-