Manufacturer's investigation conclusion: the centrimag 2nd generation primary console (serial #: (b)(6)) was returned for evaluation with the reported event of the console alarming for motor disconnect and no flow alarms and the motor speed dropping to 0 rpm.A review of the downloaded log file showed events spanning approximately 4 days (03may2022 ¿ 05may2022, 26may2022 per time stamp).Events occurring on 26may2022 took place during lab testing at abbott.The console was operating a motor at a speed of ~4600 rpm with a flow of ~5.1 lpm.On 04may1011 at 07:54 the system shutdown was initiated.A ¿motor disconnected: m2¿ alarm was active, and the speed dropped to 0 rpm and the flow was retrograde.A ¿system alert: s3¿ alarm activated.¿flow signal interrupted: f2¿ and ¿flow below minimum: f3¿ alarms were also present at these times.These alarms were able to be muted and cleared but continued to activate until the system was shutdown at 08:01.There were no other notable events active in the log file.The centrimag 2nd generation primary console was returned for analysis to the service depot.The console and returned and associated motor, and flow probe were connected to a test loop and run.During operation the motor cable was manipulated and caused the console to alarms and the pump to stop.The console and flow probe were then tested with a test motor on a test loop and operated as intended with no alarms for several days.A full functional test was performed, and all tests were passed.The reported event could not be correlated to an issue with the returned console.The device history records were reviewed for the centrimag 2nd generation primary console (serial #: (b)(6)) and the console was found to pass all manufacturing and qa specifications.The 2nd generation centrimag system operating manual section 4 entitled "warnings & precautions" warns "one additional 2nd generation centrimag primary console, motor and flow probe are required as backup system in the immediate vicinity of each patient whenever the centrimag or pedivas blood pump is used.The backup console must be connected to the backup motor and to the backup flow probe, have a battery charge sufficient for at least one hour of operation, be connected to ac power (except during transport) and be immediately available should the main console, motor or flow probe experience a malfunction." the 2nd generation centrimag system operating manual section 10 entitled "emergency and troubleshooting" states that "the recommended practice whenever there is a 2nd generation centrimag primary console or motor malfunction is to replace the console and motor as a set.Remove the blood pump from the malfunctioning motor and console and place the blood pump in the backup motor and console to continue patient support.Do not exchange individual motors or individual consoles during patient support." the 2nd generation centrimag system operating manual section 12.1 ¿appendix i ¿ console alarms and alerts¿ table 16 details how to properly interpret and troubleshoot all system alarms including m2, s3, f2, and f3 alarms.No further information was provided.The manufacturer is closing the file on this event.
|