Correction made to g1: baxter healthcare - englewood 14445 grasslands dr englewood co.80112 (previously submitted as plexus corporation 2500 millbrook drive buffalo grove, il.60089).H10: the actual device was received for evaluation.Visual inspection was performed and the top plate was observed scratched, the main module cover was stained and had corroded paint, and a scratched pump cover and worn rubber feet was observed.Unknown fluids were found on the unit.Functional tests including electrical safety and functional system level tests were performed and no issues were noted.During inspection of the power supply, burnt traces were identified on the power supply board (pcb) around components u4, u8 and capacitor c27 from fluid ingress.The reported condition was verified.The cause of the burning odor was due to the damaged power supply board, however the cause of the damage could not be determined.A service history review was performed and revealed that the device has no previous service events; therefore, servicing did not cause or contribute to the reported event.Should additional relevant information become available, a supplemental report will be submitted.
|